You can reduce the steps required to carry out the zero and calibration
NOTE
procedure as follows:
1 Connect the power sensor to the POWER REF output.
2 Press
Now set the frequency of the signal you want to measure. The power meter
automatically selects the appropriate calibration factor.
3 Press
4 Use the
Figure 4-3
5 Confirm your choice by pressing
6 Press
7 Connect the power sensor to the signal to be measured.
Keysight N1911A/1912A User's Guide
and
Zero + Cal
,
Zero + Cal
Zero + Cal A
. On dual channel meters select the required channel.
and
to display the Frequency pop-up. Use the numeric keypad to enter the
required value in the Frequency pop-up window.
Frequency pop-up
key to close the Channel Setup screen.
Now make the measurement.
The corrected measurement result is displayed.
Using E9320 E-Series Power Sensors
. (For dual channel meters, press
or
Zero + Cal B
keys to highlight the Frequency value field and press
or
MHz
as required).
.
GHz
4
171