2
General Power Meter Functions
Figure 2-4
Dependent Offset prior to any mathematical functions. These allow you to
compensate each channel individually. An overall offset can be applied if required
using the Display Offset.

Setting Channel Offsets

This gain or loss is applied to the measured power before any mathematical
functions, display offsets or relative functions are included.
Offsets are entered in dB and the range is –100 dB to +100 dB.
The equivalent range in % is from 0.0000000001% to 10000000000%.
A positive value compensates for a loss and a negative value compensates for a
gain.
Procedure
To enter a channel offset:
1 Press
2 Press
3 Use the
4 Press
Offset setting field
Figure 2-5
5 Press
62
shows that how you can apply a Channel Offset or a Frequency
to display the Channel Setup screen. Confirm the channel
requiring setup is displayed.
to display the Offsets Setup.
Offsets
and
to check the Offset setting field.
Typical channel offset display
to highlight the Offset value field and press
Offset pop-up. Use the numeric keypad to enter the required value in the
Offset pop-up window.
keys to highlight the Offset setting field.
Offset value field