3
Using P-Series Power Sensor

Setting the Trace Display

The P-Series power meter is optimized to operate in the trace display mode when
measuring a pulsed signal.
The swiftest method to get the power meter into a trace display is described in the
following procedure:
The procedure assumes you are starting from a default preset.
NOTE
1 Connect the power sensor to the power meter and connect it to the power
2 Set the Channel Frequency
3 Set the Trigger to Continuous
4 Set the Display to Trace
5 Expand window and use the Gate control and Trace control menus to set the
You can return to other menus later to add more accuracy on the measurement,
NOTE
for example, Channel Setup, to set the measurement average, the video average
and the bandwidth of a modulating signal.
6 When you are satisfied with the configuration you may want to save it for
152
source
gate markers, trigger delay, and scale.
future use. See
"Saving and Recalling Power Meter States"
further information.