Figure 2-73 Calibration Wait Pop-Up; Manual Calibration - Keysight N1911A User Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

2
General Power Meter Functions
During calibration, the wait symbol is displayed.
Figure 2-73
Offset and relative settings are ignored during calibration.

Manual calibration

To manually calibrate the power meter and sensor combination:
1 Press
For N1912A, you calibrate each channel independently by pressing the
NOTE
softkey by choosing
TIP
Press
Zero + Cal A
142
Calibration wait pop-up
and the channel
The Calibrating pop-up is then displayed.
Cal A
You can reduce the steps required to carry out the manual zero and
calibration procedure as follows:
and
Zero + Cal
or
Zero + Cal B
softkey to start the calibration routine.
Cal
or
.
Cal B
. (For dual channel meters, press
as required).
Keysight N1911A/1912A User's Guide
Cal
,
Zero + Cal