Figure 8-6 Calibration Factor Pop-Up Window - Keysight N1911A User Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

You can reduce the steps required to carry out the zero and calibration
NOTE
procedure as follows:
Now set the sensor calibration factor for the frequency of the signal you want to
measure.
9 Check the current reference calibration factor setting by pressing
10 To change the settings, use the
Figure 8-6
11 Press
12 Connect the power sensor to the signal to be measured.
13 The corrected measurement result is displayed.
Keysight N1911A/1912A User's Guide
1 Connect the power sensor to the POWER REF output.
2 Press
and
,
Zero + Cal
Zero + Cal A
. The value is displayed on the Cal Fac field.
Offset
Does this setting match the value for the sensor? (The power sensor reference
calibration factor can normally be found above the calibration factors table on
the power sensor body.)
field and press
keypad to enter the required value in the Cal Factor pop-up window.
Calibration factor pop-up window
to complete the entry.
%
Using N8480 Series Power Sensors
. (For dual channel meters, press
Zero + Cal
or
Zero + Cal B
and
to display the Cal Factor pop-up. Use the numeric
as required).
keys to highlight the Cal Fac value
8
,
239