Automatic Zero
The automatic zero can be enabled as follows:
Press
The Zeroing pop-up is displayed when a zero occurs.
When you are performing measurements, this may cause delays in obtaining
NOTE
results. During this time, it is recommended to disable the automatic zeroing.
Manual Zeroing
To manually zero the power meter and sensor:
Press
The Zeroing pop-up is displayed.
On dual channel meters you can zero each channel independently or both
NOTE
channels sequentially by pressing the
Zero Both
Calibration
Calibration sets the gain of each power meter channel and P-Series sensor
combination. This is achieved without the need to connect it to the 1.0 mW Power
Reference. The power meter can be set to automatically calibrate on-the-fly while
you are using it. Alternatively, you can perform the calibration manually. The
reference calibration factor is automatically set for all P-Series power sensors.
Keysight N1911A/1912A User's Guide
,
and toggle the
1 of 2
and the channel
Zero
General Power Meter Functions
to On.
Auto Zero A
softkey.
softkey - choosing
Zero
2
,
or
Zero A
Zero B
141