Editing/Generating sensor calibration tables

Predefined sensor calibration factor table stored in power meter is not
NOTE
applicable for Keysight N8480 Series power sensors with Option CFT. Therefore,
users are required to create a new sensor calibration table for the sensors when
a sensor calibration table is needed.
To help achieve the best accuracy in your measurement you can enter the values
supplied for the sensors you are using by editing the installed sensor calibration
tables or by generating your own custom tables.
You cannot delete any of the 20 existing calibration tables or create any additional
tables. However, you can edit or delete the content of each table. If you need
another table you should edit and re-name one of the tables.Each calibration
table can contain a maximum of 80 frequency points.
To view the calibration tables currently stored in the power meter, press
Tables
Figure
Table 8-3
Table
0
1
2
3
4
[a] DEFAULT is a sensor calibration table where the reference calibration factor and calibration factors are 100%. This
[b] The Keysight 8482B and Keysight 8482H power sensors use the same data as the Keysight 8482A.
246
,
Sensor Cal Tables
8-8.
Installed power sensor models
Sensor Model
[a]
DEFAULT
8481A
[b]
8482A
8483A
8481D
sensor calibration table can be used during the performance testing of the power meter.
. The Sensor Tbls screen is displayed as shown in
Table
5
6
7
8
9
Keysight N1911A/1912A User's Guide
Sensor Model
8485A
R8486A
Q8486A
R8486D
8487A
,