Using 8480 Series Power Sensors
Default Channel Setup
Figure 7-2
the power meter to this configuration.
Figure 7-2
8480 Series sensors connection requirements
Table 7-1
Sensor
8481A
8481H
8482A
8482H
8481D
8484A
8483A
210
shows the Channel Setup configured automatically. Presetting returns
8480 Series sensor default channel setup
8480 Series connection requirements
Connection Requirements
These power sensors connect directly to the POWER REF.
Prior to calibration, a Keysight 11708A 30 dB reference attenuator should be connected
between the power sensor and the POWER REF. Remove this attenuator from the power
sensor input before making measurements.
This power sensor requires a 75 Ω (f) to 50 Ω (m) N-Type adapter (1250-0597) to
connect to the POWER REF. Remove this adapter before making measurements.