Setting the Cable Short/Long
When connecting the power sensor with any Keysight power sensor cable with
length of 10 m (32 ft) and above, you are required to change the cable setting
from Short (default setting) to Long, or an error will occur when the power sensor
is connected.
Short
This is the default setting and for sensor cable length less than or equal to 10 m
(32 ft).
Long
This setting is for sensor cable length more than 10 m (32 ft). The maximum
supported sensor cable length is 31 m (100 ft).
To access the cable setting,
1 Press
2 In the Service menu, press
The LCB indicator will appear in blue at the bottom of the display to indicate that
the long cable option is selected as shown in
Figure 2-76
Keysight N1911A/1912A User's Guide
,
1 of 2
between the short or long cable options.
Short/long cable option
General Power Meter Functions
and select
.
Service
and select
1 of 2
Figure
to toggle
Cable Short/Long
2-76.
2
147