Procedure
Set the range as follows:
1 Press
2 Use the
3 Press
4 Use the
5 Press

Measurement considerations

While autoranging is a good starting point, it is not ideal for all measurements.
Signal conditions such as crest factor or duty cycle may cause the power meter to
select a range which is not the optimum configuration for your specific
measurement needs. Signals with average power levels close to the range switch
point require you to consider your needs for measurement accuracy and speed.
For example, using an E9300/1/4A sensor, where the range switch point is –10 ±
0.5 dBm in a pulsed signal configured as follows:
Characteristic
Peak Amplitude
Duty Cycle
The calculated average power is –12 dBm.
Keysight N1911A/1912A User's Guide
.
On dual channel meters, select the required channel.
and
keys to select the Range: setting field.
to display the Range pop-up.
and
keys to select the required setting.
to complete the entry.
Value
–6 dBm
25%
Using E9300 E-Series Power Sensors