Sensor Calibration Tables
This section describes how to use sensor calibration tables. Sensor calibration
tables store the measurement calibration factors, for a power sensor model or for
a specific power sensor, in the power meter. They are used to correct
measurement results.
Use sensor calibration tables when you want to make power measurements over a
range of frequencies using one or more power sensors.
The P-Series power meters are capable of storing 20 sensor calibration tables
each containing up to 80 frequency points. The power meter is supplied with a set
of 9 predefined sensor calibration tables plus a "100%" default table. The data in
these tables is based on statistical averages for a range of Keysight Technologies
power sensors. Your own sensor will most likely differ from the typical to some
degree. If you require best accuracy, create a custom table for each sensor you
use as shown in
To use calibration factor tables:
1 Select the sensor calibration table to be applied to a channel.
2 Zero and calibrate the power meter. The reference calibration factor used
3 Specify the frequency of the signal you want to measure. The calibration factor
4 Make the measurement.
Keysight N1911A/1912A User's Guide
"Editing/Generating sensor calibration tables"
during the calibration is automatically set by the power meter from the sensor
calibration table.
is automatically set by the power meter from the sensor calibration table.
Using 8480 Series Power Sensors
7
on page 221.
217