Using N8480 Series Power Sensors
Figure 8-7
Example
To make a measurement on channel A with a power sensor which has a reference
calibration factor of 99.8% and a calibration factor of 97.6% at the measurement
frequency.
– Disconnect the power sensor from any signal source.
– Press
– Use the numeric keypad to enter 99.8 in the Ref Cal Factor pop-up window.
– Press
– Press
– Connect the power sensor to the POWER REF output.
– Press
– Press
240
Calibration factor displayed
,
REF CFs
to complete the entry.
%
and the channel
displayed.
and the channel
The Calibrating pop-up is then displayed.
,
Offset
and the channel
softkey. The Zeroing pop-up is
Zero
softkey to start the calibration routine.
Cal
. The value is displayed on the Cal Fac field.
Keysight N1911A/1912A User's Guide
Calibration factor
softkey.
REF CF