Figure 2-40 Trigger Setting Menu 2 Of 2 - Keysight N1911A User Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

2
General Power Meter Functions
Figure 2-40
Slope
The current setting is displayed below the
displayed in single numeric display mode.
trigger from an increasing power level. Similarly,
the trigger from a decreasing power level. To change the setting, press
highlight
Holdoff
The current setting is displayed below the
occurs, the trigger mechanism is disabled for the configured time period. This
allows stable triggering to be achieved even when a signal has multiple edges, for
example, a TDMA signal with non-constant amplitude modulation. Values up to
400 ms can be configured.
To change the setting press
required value in the Trigger Holdoff pop-up window to enter the new value.
100
Trigger setting menu 2 of 2
or
as required.
+
Holdoff
label and the
Slope
(and
) is used to generate the
+
(and
label. After a trigger event
Holdoff
and use the numeric keypad to enter the
Keysight N1911A/1912A User's Guide
or
symbol is
) is used to generate
to
Slope