Sensor Calibration Tables - Keysight N1913 User Manual

Epm series power meters
Hide thumbs Also See for N1913:
Table of Contents

Advertisement

Sensor Calibration Tables

This section describes how to use sensor calibration tables. Sensor calibration
tables store the measurement calibration factors, for a power sensor model or for
a specific power sensor, in the power meter. They are used to correct
measurement results.
Use sensor calibration tables when you want to make power measurements over a
range of frequencies using one or more power sensors.
The N1913/1914A EPM Series power meters are capable of storing 20 sensor
calibration tables, each containing up to 80 frequency points. The power meter is
supplied with a set of nine predefined sensor calibration tables plus a "100%"
default table. The data in these tables is based on statistical averages for a range
of Keysight Technologies power sensors. Your own sensor will most likely differ
from the typical to some degree. If you require best accuracy, create a custom
table for each sensor you use as shown in
Tables"
To use calibration factor tables,
1 Select the sensor calibration table to be applied to a channel.
2 Zero and calibrate the power meter. The reference calibration factor used
3 Specify the frequency of the signal you want to measure. The calibration factor
4 Make the measurement.
Keysight N1913/1914A User's Guide
on page 191.
during the calibration is automatically set by the power meter from the sensor
calibration table.
is automatically set by the power meter from the sensor calibration table.
Using N8480 Series Power Sensors
"Editing/Generating Sensor Calibration
6
187

Advertisement

Table of Contents
loading

This manual is also suitable for:

1914a

Table of Contents