Figure 6-6 Calibration Factor Pop-Up Window - Keysight N1913 User Manual

Epm series power meters
Hide thumbs Also See for N1913:
Table of Contents

Advertisement

6
Using N8480 Series Power Sensors
You can reduce the steps required to carry out the zero and calibration
NOTE
procedure as follows:
– Connect the power sensor to the POWER REF output.
– Press
Now, set the sensor calibration factor for the frequency of the signal to measure.
9 Check the current reference calibration factor setting by pressing
Does this setting match the value for the sensor?
NOTE
The power sensor reference calibration factor can normally be found above the
calibration factors table on the power sensor body.
10 To change the settings, use the
Figure 6-6
184
and
Zero + Cal
or
Zero + Cal A
Zero + Cal B
. The value is displayed on the Cal Fac field.
Offset
value field and press
keypad to enter the required value in the Cal Factor pop-up window.
Calibration factor pop-up window
. (For dual channel meters, press
as required).
and
keys to highlight the Cal Fac
to display the Cal Factor pop-up. Use the numeric
Keysight N1913/1914A User's Guide
,