Adapter Test Procedure - Keysight N1913 User Manual

Epm series power meters
Hide thumbs Also See for N1913:
Table of Contents

Advertisement

Adapter Test Procedure

System Calibration
Table 2-6
Equipment list
Instrument
Performance network
analyzer (PNA)
Adapters
Adapters
Mechanical
calibration kit
Mechanical
calibration kit
Mechanical
calibration kit
11500E cable assembly,
3.5 mm (m)
to 3.5 mm (m)
Table 2-7
Typical specifications
Frequency
50 MHz
Keysight N1913/1914A User's Guide
Critical specifications
10 MHz to 50 GHz
2.4 mm (f), Type-N (f)
2.4 mm (f), 3.5 mm (m)
Type-N, 50 Ω
3.5 mm, 50 Ω
2.4 mm, 50 Ω
3.5 mm (m) to 3.5 mm
(m), DC to 26.5 GHz
S-Parameter
S11 and S22
S12 and S21
0.999235 dB (lin mag) <0.1%
General Power Meter Functions
Recommended model
E8362B/C, E8363B/C,
N5225A, or similar
11903B
11901D
85054B
85052B
85056A
11500E
08485-60005
−58 dB
−0.00665 dB (log mag)
Manufacturer
Keysight Technologies
Keysight Technologies
Keysight Technologies
Keysight Technologies
Keysight Technologies
Keysight Technologies
Keysight Technologies
08487-60001
−53 dB
−0.006908 dB (log mag)
0.999205 dB (lin mag)
<0.1%