Measurement Accuracy - Keysight N1913 User Manual

Epm series power meters
Hide thumbs Also See for N1913:
Table of Contents

Advertisement

Measurement Accuracy

Power sensors have small errors in their response over frequency. The response of
each sensor is measured during manufacture (and during periodic calibration).
With E-Series power sensors, the resulting frequency compensation information is
written into Electrically Erasable Programmable Read Only Memory (EEPROM).
This allows the frequency and calibration data to be downloaded to the power
meter automatically.
Using calibration factors enables you to achieve improved measurement
accuracy. This section describes making continuous wave measurements using
the E4410 E-Series power sensors.
Making a measurement requires the following steps:
1 Zero and calibrate the power meter/sensor combination.
2 Set the frequency for the signal you want to measure.
3 Make the measurement.
Procedure
1 Zero and calibrate the power meter/sensor combination.
2 Ensure the power sensor is disconnected from any signal source.
3 Press
4 Connect the power sensor to the POWER REF output.
5 Press
Keysight N1913/1914A User's Guide
and the channel
and the channel
Calibrating pop-up is then displayed.
Using E4410 E-Series Power Sensors
softkey. The Zeroing pop-up is displayed.
Zero
softkey to start the calibration routine. The
Cal
4
151

Advertisement

Table of Contents
loading

This manual is also suitable for:

1914a

Table of Contents