Figure 2-9 Frequency Dependent Offset Indicator - Keysight N1913 User Manual

Epm series power meters
Hide thumbs Also See for N1913:
Table of Contents

Advertisement

When no data is contained in the highlighted table, the
NOTE
(grayed out).
3 Press
4 Press
Offset Table A Selected
Figure 2-9
5 To change the frequency, press
6 Press
7 To confirm your choice, press the appropriate unit softkey.
8 Connect the power sensor to the signal to be measured.
9 The measurement result, including offset, is now displayed.
If the measurement frequency does not correspond directly to a frequency in the
NOTE
sensor calibration table (if selected) and the frequency dependent offset table
being used, the power meter calculates the calibration factor and offset using
linear interpolation. If you enter a frequency outside the frequency range defined
in the sensor calibration table or the frequency dependent offset table, the
power meter uses the highest or lowest frequency point in the appropriate table
to set the calibration factor and offset.
Keysight N1913/1914A User's Guide
to complete the selection of the offset table.
again to display the measurement screen.
which offset table is selected.
Frequency dependent offset indicator
highlight the Frequency field.
to display the Frequency pop-up window. Use the numeric
keypad to enter the required value in the Frequency pop-up window.
General Power Meter Functions
Table
Figure 2-9
and use the
and
2
key is disabled
shows
keys to
67