This section describes how to zero and calibrate the power meter. You should
always zero the power meter prior to calibrating it.

Zeroing the Power Meter

Zeroing adjusts the power meter for a zero power reading with no power applied
to the power sensor. During zeroing, which takes approximately 10 seconds, the
wait symbol is displayed.
Figure 2-41
When to Zero?
Zeroing of the power meter is recommended:
– When a 5 oC change in temperature occurs.
– Every 24 hours.
– Prior to measuring low level signals. For example, 10 dB above the lowest
To zero the power meter:
– Press
– The Zeroing pop-up is displayed.
On dual channel meters you can zero each channel independently or both
NOTE
channels sequentially by pressing the
Zero Both
102
Zeroing pop-up
specified power for your power sensor.
and the channel
softkey.
Zero
softkey - choosing
Zero