Figure 5-3 Reference Calibration Factor Pop-Up Window - Keysight N1913 User Manual

Epm series power meters
Hide thumbs Also See for N1913:
Table of Contents

Advertisement

5
Using 8480 Series Power Sensors
Figure 5-3
5 Press
Now, zero and calibrate the power meter/sensor combination as follows:
6 Press
7 Connect the power sensor to the POWER REF output.
8 Press
You can reduce the steps required to carry out the zero and calibration
NOTE
procedure as follows:
– Connect the power sensor to the POWER REF output.
– Press
Now, set the sensor calibration factor for the frequency of the signal you want to
measure.
9 Check the current reference calibration factor setting by pressing
Does this setting match the value for the sensor? (The power sensor reference
NOTE
calibration factor can normally be found above the calibration factors table on
the power sensor body.)
160
Reference Calibration Factor pop-up window
to complete the entry.
%
and the channel
and the channel
Calibrating pop-up is then displayed.
and
Zero + Cal
,
Zero + Cal
Zero + Cal A
. The value is displayed on the Cal Fac field.
Offset
softkey. The Zeroing pop-up is displayed.
Zero
softkey to start the calibration routine. The
Cal
. (For dual channel meters, press
or
as required).
Zero + Cal B
Keysight N1913/1914A User's Guide
,