Keysight N1913 User Manual page 230

Epm series power meters
Hide thumbs Also See for N1913:
Table of Contents

Advertisement

9
Maintenance
Data questionable;Lower window log error
231
This indicates that a difference measurement in the lower window has given a negative result when the
units of measurement were logarithmic.
Data questionable;Upper window log error
231
This indicates that a difference measurement in the upper window has given a negative result when the
units of measurement were logarithmic.
Data questionable;ZERO ERROR
231
Power meter zeroing failed. The most likely cause is attempting to zero when some power signal is being
applied to the power sensor.
Data questionable;ZERO ERROR ChA
231
Power meter zeroing failed on Channel A. The most likely cause is attempting to zero when some power
signal is being applied to the power sensor.
Data questionable;ZERO ERROR ChB
231
Power meter zeroing failed on Channel B. The most likely cause is attempting to zero when some power
signal is being applied to the power sensor.
Hardware missing
241
The power meter is unable to execute the command because either no power sensor is connected or it
expects an E-series power sensor and one is not connected.
System error;Dty Cyc may impair accuracy with ECP sensor
310
This indicates that the sensor connected is for use with CW signals only.
System error;Ch A Dty Cyc may impair accuracy with ECP sensor
310
This indicates that the sensor connected to Channel A is for use with CW signals only.
System error;Ch B Dty Cyc may impair accuracy with ECP sensor
310
This indicates that the sensor connected to Channel B is for use with CW signals only.
System error;Sensor EEPROM Read Failed - critical data not found or unreadable
310
This indicates a failure with your E-Series Power Sensor. Refer to your power sensor manual for details on
returning it for repair.
System error;Sensor EEPROM Read Completed OK but optional data block(s) not found or unreadable
310
This indicates a failure with your E-Series Power Sensor. Refer to your power sensor manual for details on
returning it for repair.
System error;Sensor EEPROM Read Failed - unknown EEPROM table format
310
This indicates a failure with your E-Series Power Sensor. Refer to your power sensor manual for details on
returning it for repair.
230
Keysight N1913/1914A User's Guide