Programming Language Selection (Option 200)

The programming language selection is available as an orderable option. For
N1913A, you can use SCPI, HP 436A, or HP 437B programming language to
program the power meter from the remote interface. For N1914A, you can use
either SCPI or HP 438A programming language to program the power meter from
the remote interface. The default language is SCPI when the power meter is
shipped from the factory.
The power meter complies with the rules and regulations of the 1996.0 version of
SCPI (Standard Commands for Programmable Instruments). You can determine
the SCPI version with which the power meter is in compliance by sending the
SYSTem:VERSion? command from the remote interface. You cannot query the
SCPI version from the front panel.
The language selection is stored in non-volatile memory and does not change
when power has been off or after a remote interface reset.
To select the interface language from the front panel (N1913A),
1 Press
2 Select the language from HP 436A
To select the interface language from the front panel (N1914A),
1 Press
2 Select the language from HP 438A
To select the interface language from the remote interface, use the
SYSTem:LANGuage command.
Option 200 can only support 8480 series, N8480 series CFT-option, and E4412/
NOTE
3A power sensors.
[1] HP 436A, HP 437B, and HP 438A language modes are not compatible with LAN or USBTM remote
[2] SCPI is the default factory setting. License is needed for the other languages to be available.
Keysight N1913/1914A User's Guide
, Remote Interfaces, 1 of 2, and Command Set.
, Remote Interfaces, 1 of 2, and Command Set.
interface.
[1]
[1]
, HP 437B
, and SCPI
[1]
[2]
, and SCPI
.
Introduction
1
[2]
.
53