Warm Start

The warm start feature allows you to retain the meter's same states and settings
upon power cycle or in the event of interrupted power. The warm start ON/OFF is
selectable through the front panel and SCPI. The default setting is ON.
Warm start ON
All the states entered through the front panel or remote interface will be saved.
The power meter will be powered on with the states before it was powered off.
Warm start OFF
The power meter will be powered on with the default states setting.
To enable or disable the warm start feature,
1 Press
2 In the Service menu, press
Figure 2-61
Keysight N1913/1914A User's Guide
,
1 of 2
start feature. This feature is set to ON by default (see
Enable/disable warm start feature
General Power Meter Functions
and select
.
Service
Warm Start Off/On
to enable or disable the warm
Figure
2-61).