Keysight N1913 User Manual page 162

Epm series power meters
Hide thumbs Also See for N1913:
Table of Contents

Advertisement

5
Using 8480 Series Power Sensors
Example
To make a measurement on channel A with a power sensor which has a reference
calibration factor of 99.8% and a calibration factor of 97.6% at the measurement
frequency.
1 Disconnect the power sensor from any signal source.
2 Press
3 Use the numeric keypad to enter 99.8 in the Ref Cal Factor pop-up window.
4 Press
5 Press
6 Connect the power sensor to the POWER REF output.
7 Press
8 Press
9 Use the
10 Press
11 Connect the power sensor to the signal to be measured.
12 The corrected measurement result is displayed.
When no sensor tables are selected and
NOTE
chosen, the calibration factor used for the measurement is displayed in the
upper window as shown in
162
,
REF CFs
to complete the entry.
%
and the channel
and the channel
Calibrating pop-up is then displayed.
,
Offset
and
to display the Cal Factor pop-up. Use the numeric keypad to enter 97.6 in the
Cal Factor pop-up window.
to complete the entry.
%
and the channel
softkey. The Zeroing pop-up is displayed.
Zero
softkey to start the calibration routine. The
Cal
. The value is displayed on the Cal Fac field.
keys to highlight the Cal Fac value field and press
Single Numeric
Figure
5-5.
softkey.
REF CF
display mode is
Keysight N1913/1914A User's Guide