Sensor Calibration Tables

This section describes how to use sensor calibration tables. Sensor calibration
tables store the measurement calibration factors, for a power sensor model or for
a specific power sensor, in the power meter. They are used to correct
measurement results.
Use sensor calibration tables when you want to make power measurements over a
range of frequencies using one or more power sensors.
The N1913/1914A EPM Series power meters are capable of storing 20 sensor
calibration tables each containing up to 80 frequency points. The power meter is
supplied with a set of 9 predefined sensor calibration tables plus a "100%" default
table. The data in these tables is based on statistical averages for a range of
Keysight Technologies power sensors. Your own sensor will most likely differ from
the typical to some degree. If you require best accuracy, create a custom table for
each sensor you use as shown in
on page 166.
To use calibration factor tables:
1 Select the sensor calibration table to be applied to a channel.
2 Zero and calibrate the power meter. The reference calibration factor used
3 Specify the frequency of the signal you want to measure. The calibration factor
4 Make the measurement.
Keysight N1913/1914A User's Guide
during the calibration is automatically set by the power meter from the sensor
calibration table.
is automatically set by the power meter from the sensor calibration table.
Using 8480 Series Power Sensors
"Editing/Generating Sensor Calibration Tables"
5
163