Keysight N1913 User Manual page 108

Epm series power meters
Hide thumbs Also See for N1913:
Table of Contents

Advertisement

2
General Power Meter Functions
Sensor Model
Keysight 8485D
Keysight 8487A
Keysight 8487D
108
Connection Requirements
Prior to the power meter being zeroed and calibrated, a Keysight 11708A 30 dB reference
attenuator and an APC 3.5 (f) to 50 Ω (m) N-Type adapter (08485-60005) should be
connected between the power sensor and the reference calibrator. Remove this attenuator
from the power sensor input before making measurements.
This power sensor requires an APC 2.4 (f) to 50 Ω (m) N-Type adapter (08487-60001) to
connect to the power meter. Remove this attenuator before making measurements.
Prior to the power meter being zeroed and calibrated, a Keysight 11708A 30 dB reference
attenuator and an APC 2.4 (f) to 50 Ω (m) N-Type adapter (08487-60001) should be
connected between the power sensor and the reference calibrator. Remove this attenuator
from the power sensor input before making measurements.
Keysight N1913/1914A User's Guide