6 Confirm your choice by pressing
7 Press
Figure 2-6
The Ofs symbol is not displayed when the associated measurement is displayed
NOTE
in Dual Numeric or Analog format.

Setting Display Offsets

This gain or loss is applied to the measured power after any channel offsets or
mathematical functions have been included.
Offsets are entered in dB and the range of values is –100 dB to +100 dB. A positive
value compensates for a loss, and a negative value compensates for a gain.
Procedure
Enter a display offset on the currently selected window:
1 Press
2 Select the window you wish to set the offset value on by pressing the
3 Use the
4 Press
Keysight N1913/1914A User's Guide
key to complete the offset entry. If either a channel or a display
offset is set, the Ofs indicator is displayed.
Channel offset indicator
to display the Measurement Setup screen.
key. The currently selected window/measurement is displayed.
Meas Select
and
to check the Offset setting field.
General Power Meter Functions
.
dB
keys to highlight the Offset setting field.
Offset indicator
2