10 Press
11 Set the frequency range and step by sending the below commands to meter
Refer to the EPM Series Power Meters Programming Guide for the commands
NOTE
usage details.
Example of commands set:
FREQ:STAR 10MHz
FREQ:STOP 500MHz
FREQ:STEP 10
The power meter can be remotely controlled by LAN, USB and GPIB (IEEE488)
NOTE
programming interfaces. Refer to N1913/1914A EPM Series Power Meters
Installation Guide for details.
12 Send *OPC (OPeration Complete) command to meter; to set the operation
13 Set the meter to continuous trigger cycles by sending INITiate:CONTinuous
14 Configure and set the required power sweep range and step of the power
15 Set the power source trigger input and trigger output.
16 Set the power source to sweep.
Keysight N1913/1914A User's Guide
or
measurement averaging, measurement frequency, offsets, duty cycle and so
forth. Refer to
"General Power Meter Functions"
procedures.
using remote interface.
– SENSe1:FREQuency:STARt <start_frequency><frequency_unit>
– SENSe1:FREQuency:STOP <stop_frequency><frequency_unit>
– SENSe1:FREQuency:STEP <frequency_step_size>
complete bit in the Standard Event Status register when all pending device
operations are completed.
ON command to meter.
source.
General Power Meter Functions
to setup measurement settings such as
on page 55 for the setup
2