Figure 6-9 Frequency Dependent Offset Indicator - Keysight N1913 User Manual

Epm series power meters
Hide thumbs Also See for N1913:
Table of Contents

Advertisement

When no data is contained in the highlighted table, the
NOTE
(grayed out).
3 Press
4 Press
Sensor Table 2 Selected
Figure 6-9
5 To change the frequency, press
6 Press
7 To confirm your choice, press the appropriate unit softkey.
8 Connect the power sensor to the signal to be measured.
9 The corrected measurement result is now displayed.
If the measurement frequency does not correspond directly to a frequency in the
NOTE
sensor calibration table, the power meter calculates the calibration factor using
linear interpolation.
If you enter a frequency outside the frequency range defined in the sensor
calibration table, the power meter uses the highest or lowest frequency point in
the sensor calibration table to set the calibration factor.
Keysight N1913/1914A User's Guide
to complete the selection of the calibration factor table.
Done
again to display the measurement screen.
Done
offset table is selected.
Frequency dependent offset indicator
highlight the Frequency field.
to display the Frequency pop-up window. Use the numeric
keypad to enter the required value in the Frequency pop-up window.
Using N8480 Series Power Sensors
Table
Figure 6-9
and use the
key is disabled
shows which
and
keys to
189