Power Sensor Connection Requirements - Keysight N1913 User Manual

Epm series power meters
Hide thumbs Also See for N1913:
Table of Contents

Advertisement

Table 2-5
Sensor Model
Keysight 8481A
Keysight 8481H
Keysight 8482A
Keysight 8482H
Keysight N8481A
Keysight N8482A
Keysight E4412A
Keysight E930xA
Keysight E930xH
Keysight E9304 H18
Keysight E9304 H19
Keysight 8481D
Keysight 8484A
Keysight 8483A
Keysight R8486A
Keysight Q8486A
Keysight V8486A
Keysight W8486A
Keysight E8486A
Keysight R8486D
Keysight Q8486D
Keysight 8481B
Keysight 8482B
Keysight E930xB
Keysight 8485A
Keysight N8485A
Keysight E4413A
Keysight E9300A H24
Keysight E9300A H25
Keysight N1913/1914A User's Guide

Power Sensor Connection Requirements

Connection Requirements
These power sensors connect directly to the reference calibrator.
Prior to the power meter being calibrated, a Keysight 11708A 30 dB reference attenuator
should be connected between the power sensor and the reference calibrator. Remove this
attenuator from the power sensor input before making measurements.
This power sensor requires a 75 Ω (f) to 50 Ω (m) N-Type adapter (1250-0597) to connect
to the reference calibrator. Remove this adapter from the power sensor input before
making measurements.
The waveguide power sensors have two connectors. The N-Type connector is the one
which is used to calibrate the power meter.
These power sensors are configured with an attenuator. Prior to the power meter being
calibrated, this attenuator must be removed. The attenuator must be reconnected prior to
making measurements.
This power sensor requires an APC 3.5 (f) to 50 Ω (m) N-Type adapter (08485-60005) to
connect to the reference calibrator. Remove this adapter before making measurements.
General Power Meter Functions
2
107