Figure 2-36 Recorder Maximum Pop-Up - Keysight N1913 User Manual

Epm series power meters
Hide thumbs Also See for N1913:
Table of Contents

Advertisement

Figure 2-36
11 Use the numeric keys to enter the power level you want to generate a 0 Vdc
12 Press
The recorder output can be disabled and re-enabled by checking the Rec o/p:
NOTE
setting field.
The highest power you are going to measure is used to determine the value
which you should set for the Recorder Output maximum setting. For example, if
you are measuring a power less than 1 mW and greater than 100 mW, then set
the recorder maximum value to 1 mW.
Table 2-4
Log
50
Lin
100 W
Log
–10
Lin
100 mW
Keysight N1913/1914A User's Guide
Recorder Maximum pop-up
output in the Recorder Minimum pop-up and press
key to close the Measurement Setup screen.
Ranges of recorder output setting
40
10 W
–20
10 mW
General Power Meter Functions
30
20
1 W
100 mW
–30
–40
1 mW
100 nW
.
dBm
10
0
10 mW
1 mW
–50
–60
10 nW
1 nW
2
99

Advertisement

Table of Contents
loading

This manual is also suitable for:

1914a

Table of Contents