Frequency Sweep Mode - Keysight N1913 User Manual

Epm series power meters
Hide thumbs Also See for N1913:
Table of Contents

Advertisement

2
General Power Meter Functions

Frequency Sweep Mode

Frequency sweep is generally used in a frequency response calibration system
where the amplitude is fixed, and the frequency of the power source signal is
swept. This mode can be used to determine the frequency response of a device
under test.
This feature is only available when N8480 Series, E-Series E4410, or E-Series
NOTE
E9300 sensor is connected.
Procedures
1 Connect sensor to a power source.
2 Connect power meter TRIG OUT to power source TRIG IN using a BNC cable.
3 Press
When N8480 Series, E-Series E4410, E-Series E9300 or U2000 Series sensor is
NOTE
connected, Sensor Mode is set to AVG only by default.
4 Press
5 Press
6 Select either
7 Press
8 Press
9 Press
84
The same connection is applied for power source TRIG OUT to power meter
TRIG IN. See
Figure
. The Channel Setup displays as shown in
. The Trigger menu is displayed.
Acqn
softkey to configure a trigger.
Sing Trig
Settings
to configure the remaining trigger parameters. The trigger
Settings menu consists of two pages.
Figure 2-22
shows page 2.
and
Source
Ext
,
1 of 2
Output
rear panel TRIG OUT BNC port when the meter is triggered upon a completed
measurement.
2-19.
or
.
Cont Trig
Figure 2-21
will be automatically enabled.
and
to allow a TTL level high to be produced at the
On
Figure
shows page 1 and
Keysight N1913/1914A User's Guide
2-20.