Using E9300 E-Series Power Sensors
Procedure
1 Zero and calibrate the power meter/sensor combination.
2 Ensure the power sensor is disconnected from any signal source.
3 Press
4 Connect the power sensor to the POWER REF output.
5 Press
You can reduce the steps required to carry out the zero and calibration
NOTE
procedure as follows:
– Connect the power sensor to the POWER REF output.
– Press
Now set the frequency of the signal you want to measure. The power meter
automatically selects the appropriate calibration factor.
6 Press
7 Use the
136
and the channel
and the channel
Calibrating pop-up is then displayed.
and
Zero + Cal
,
Zero + Cal
Zero + Cal A
. On dual channel meters select the required channel.
and
to display the Frequency pop-up. Use the numeric keypad to enter the
required value in the Frequency pop-up window.
softkey. The Zeroing pop-up is displayed.
Zero
softkey to start the calibration routine. The
Cal
. (For dual channel meters, press
or
Zero + Cal B
keys to highlight the Frequency value field and press
as required).
Keysight N1913/1914A User's Guide