Refer to the N1913/1914A EPM Series Power Meters Programming Guide for the
NOTE
commands usage details.
Example of command set:
BUFF:COUN 100
The power meter can be remotely controlled by LAN, USB and GPIB (IEEE488)
NOTE
programming interfaces. Refer to N1913/1914A EPM Series Power Meters
Installation Guide for details.
12 Press
13 Send *OPC (OPeration Complete) command to meter to set the operation
14 Set the meter to continuous trigger mode by sending INITiate:CONTinuous
15 Configure and set the required power sweep range and step on the power
16 Set the power source trigger input and trigger output, then start sweeping.
17 Poll the status of the power meter by sending *ESR?. *ESR? will return a 1
Keysight N1913/1914A User's Guide
or
averaging, measurement frequency, offsets, duty cycle and so forth. Refer to
"General Power Meter Functions"
complete bit in the Standard Event Status register when all pending device
operations are completed.
ON command to meter.
source accordingly.
when buffering is completed. Use FETCh? to retrieve all the buffered
measurement.
General Power Meter Functions
to setup measurement setting such as measurement
on page 55 for the setup procedures.