Measurement Accuracy And Speed; Setting The Range - Keysight N1913 User Manual

Epm series power meters
Hide thumbs Also See for N1913:
Table of Contents

Advertisement

Measurement Accuracy and Speed

The power meter has no internal ranges. The only ranges you can set are those of
the U8480 Series USB thermocouple sensors. With an U2000 Series power sensor
the range can be set either automatically or manually. Use autoranging when you
are unsure of the power level you are about to measure.
To prevent damage to your sensor do not exceed the power levels specified
CAUTION
in the sensor user's guide. The U2004A sensor is DC coupled. DC voltages in
excess of the maximum value (5 VDC) can damage the sensing diode.

Setting the Range

There are two manual settings, LOWER and UPPER. The LOWER range uses the
more sensitive path and the UPPER range uses the attenuated path in the U8480
Series USB thermocouple sensors.
Sensor
U2000A, U2001A, U2002A, U2004A
U2000H, U2001H, U2002H
U2000B, U2001B
The default is AUTO. In AUTO the range crossover value depends on the sensor
model being used.
U2000A, U2001A, U2002A, U2004A
U2000H, U2001H, U2002H
U2000B, U2001B
Keysight N1913/1914A User's Guide
Sensor
Using U2000 Series USB Power Sensors
LOWER range
60 dBm to
10 dBm
–50 dBm to 0 dBm
–30 dBm to +20 dBm
Range Crossover Values
UPPER range
10 dBm to +20 dBm
0 dBm to +30 dBm
+20 dBm to +44 dBm
10 dBm + 1 dB
0 dBm + 1 dB
+20
dBm + 1 dB
7
203

Advertisement

Table of Contents
loading

This manual is also suitable for:

1914a