Using N8480 Series Power Sensors
6 Connect the power sensor to the POWER REF output.
7 Press
8 Press
9 Use the
10 Press
11 Connect the power sensor to the signal to be measured.
12 The corrected measurement result is displayed.
When no sensor tables are selected and
NOTE
chosen, the calibration factor used for the measurement is displayed in the
upper window as shown in
186
and the channel
Calibrating pop-up is then displayed.
,
Offset
and
to display the Cal Factor pop-up. Use the numeric keypad to enter 97.6 in the
Cal Factor pop-up window.
to complete the entry.
%
softkey to start the calibration routine. The
Cal
. The value is displayed on the Cal Fac field.
keys to highlight the Cal Fac value field and press
Single Numeric
Figure
6-7.
display mode is
Keysight N1913/1914A User's Guide