To make a measurement on channel A with a power sensor which has a reference
calibration factor of 99.8% and a calibration factor of 97.6% at the measurement
frequency.
1 Disconnect the power sensor from any signal source.
2 Press
3 Use the numeric keypad to enter 99.8 in the Ref Cal Factor pop-up window.
4 Press
5 Press
Keysight N1913/1914A User's Guide
to complete the entry.
%
Calibration factor displayed
,
and the channel
REF CFs
to complete the entry.
%
and the channel
Using N8480 Series Power Sensors
softkey.
REF CF
softkey. The Zeroing pop-up is displayed.
Zero
Calibration factor
6
185