Figure 2-44 System Calibration Setup - Keysight N1913 User Manual

Epm series power meters
Hide thumbs Also See for N1913:
Table of Contents

Advertisement

2
General Power Meter Functions
Figure 2-44
1 Set up the equipment for system calibration according to
2 Preset all test equipment.
3 Perform the following settings on the PNA:
4 Extend Port 1 of the PNA by connecting it to a 2.4 mm (f) to Type-N (f)
5 Extend Port 2 of the PNA by connecting it to a 2.4 mm (f) to 3.5 mm (f) adapter
6 Perform one port calibration (at Port 1) using the 85054B Mechanical Cal Kit
7 Perform one port calibration (at Port 2) using the 85052B/85056A Mechanical
8 When the wizard prompts for a through adapter connection, attach the
9 The firmware will estimate the delay, which should be approximately 0.11 ns.
110
PNA
Port 1
PNA
Port 2
System calibration setup
– Start frequency = 10 MHz
– Stop frequency = 100 MHz
– Number of points = 19
– Power = –8 dBm
– IF bandwidth = 10 Hz
– Averaging = 1
– Sweep time = 1 s
– Sweep auto = true
– Smoothing = 0%
adapter.
via a semi-rigid cable.
– If you are verifying the 08487-60001, connect a 3.5 mm (f) to 2.4 mm (m)
adapter to a semi-rigid cable.
(Open, Short, Load, and Sliding Load).
Cal Kit (Open, Short, and Load).
adapter-under-test as the unknown adapter.
Open, Short, Load, Sliding Load
Open, Short, Load, Sliding Load
Keysight N1913/1914A User's Guide
Figure
2-44.