Calibration Procedure Using E-Series Power Sensors and N8480
Series Power Sensors (excluding Option CFT)
The following procedure describes how you calibrate the power meter with an E-
Series power sensor or N8480 Series power sensor (excluding Option CFT). Since
the power meter automatically downloads the E- Series power sensor or N8480
Series power sensor (excluding Option CFT)'s calibration table, there is no
requirement to enter the reference calibration factor. The power meter identifies
that an E- Series power sensor or N8480 Series power sensor (excluding Option
CFT) is connected and will not allow you to select certain softkeys. The text on
these softkeys appears grayed out.
To calibrate:
1 Press
2 Connect the power sensor to the POWER REF output.
3 Press the
Keysight N1913/1914A User's Guide
.
softkey to start the calibration routine. The Calibrating pop-up
Cal
is then displayed (the power meter automatically turns on the POWER REF
output).