9
Maintenance
Query UNTERMINATED
420
The power meter was addressed to talk (that is, to send data over the interface) but a command has not
been received which sends data to the output buffer. For example you may have executed a CONFigure
command (which does not generate data) and then attempted to read data from the remote interface.
Query DEADLOCKED
430
A command was received which generates too much data to fit in the output buffer and the input buffer is
also full. Command execution continues but data is lost.
Query UNTERMINATED after indefinite response
440
The *IDN? command must be the last query command within a command string.