  • Page 1 Keysight Technologies N9030B Signal Analyzer Option 107 Audio Input and Digitizer Upgrade Kit Installation Note...
  • Page 2 The information contained in this document is subject to change without notice. Keysight Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
  • Page 3 Information on how to obtain this software can be found at: www.keysight.com/find/calibrationsoftware While Keysight does recommend that a full calibration be performed after the installation of this upgrade, the end user must ultimately determine whether they want this service or not. If a full calibration is required, arrangements regarding the level of calibration must be made between the end user and the calibration provider.
  • Page 4 This kit provides all parts and instructions necessary to install Option 107, Audio Input and Digitizer, into the Keysight N9030B. 1. This upgrade kit is not compatible with N9030B PXA signal analyzers with any of the following options currently installed:...
  • Page 5: Installation Kit Parts List

    Option 107, Audio Input and Digitizer Upgrade Kit Installation Kit Parts List Quantity Description Keysight Part Number Screw, FLT-HD Torx T10 M3 x 0.5 (6 mm long) 0515-1227 Lock Washer, Internal tooth 2190-0068 Hex Nut 15/32-32 Interface Board 2950-0035 Coaxial Cable Assembly w/Toroids...
  • Page 6: Tools Required

    Option 107, Audio Input and Digitizer Upgrade Kit Tools Required — T-10 Torx Driver — T-20 Torx Driver — 5/16-inch torque wrench — 1/4-inch open end wrench — 1/4-inch socket on 4 lb torque wrench — scissors or knife — diagonal cutters —...
  • Page 7: Installation Procedure

    Option 107, Audio Input and Digitizer Upgrade Kit Installation Procedure If the instrument is placed on its face during any of the following procedures, be sure to use a soft surface or soft cloth to avoid damage to the front panel, keys, or input connector. Make sure any adapters on the front panel are removed.
  • Page 8 Option 107, Audio Input and Digitizer Upgrade Kit Figure 1 Instrument Outer Case Removal Installation Note N9030-90111...
  • Page 9 Option 107, Audio Input and Digitizer Upgrade Kit Top Brace Removal 1. Refer to Figure 2. To remove the top brace (1), use the T-10 driver to remove the twelve panhead screws (3) (0515-0372) attaching the top brace to the chassis. Remove the twelve screws (fourteen screws with Option B1X) (4) (0515-1946) attaching the top brace to the boards.
  • Page 10 Option 107, Audio Input and Digitizer Upgrade Kit Option 107 Hardware Installation 1. Refer to Figure 3. Locate and install the N9030-60047 Audio Digitizer PCA into slot 7. Figure 3 Option 107 Audio Digitizer Installation Note N9030-90111...
  • Page 11 Option 107, Audio Input and Digitizer Upgrade Kit Audio Input Connector Installation 1. Remove the overlay to the left of the front panel display by peeling it off. 2. Refer to Figure 4.Using the T-10 driver, remove the two screws (1) (0515-1227) securing the cover plate that was behind the front panel overlay.
  • Page 12 Option 107, Audio Input and Digitizer Upgrade Kit 7. Refer to Figure 5. Install the lock washer (1) (2190-0068) and hex nut (2) (2950-0035) onto the BNC connector and tighten to 21 in-lbs. Figure 5 BNC Connector 8. Locate the new Audio Input Overlay provided with this kit (N9030-80062). 9.
  • Page 13: Final Installation

    4. Replace the four instrument bottom feet. 5. Replace the four key locks to the bottom feet. Install Instrument Software http://www.keysight.com/find/pxa_software This will guarantee: — The analyzer has the latest instrument software version to support Option 107. — FPGA updates are done to the Audio Digitizer hardware.
  • Page 14 USB Storage Device Configuration Menu 7. The PXA will automatically consume the License File (this may take a few minutes). When the License File is consumed the Keysight License Manager will display a “Successful License Installation” message similar to the one shown in...
  • Page 15: Verify The Installation

    Hardware on the analyzer and verify that the Audio Digitizer identifies itself as: — Audio Input 100 K Ohms 3. Press System, Show System on the analyzer to display a list of all installed options. You should see N9030B-107. Installation Note N9030-90111...
  • Page 16: Adjustments And Performance Verification

    If you require further assistance, please contact the Keysight support team. Online assistance: http://www.keysight.com/find/assist If you do not have access to the Internet, contact your local Keysight Technologies Sales and Service Office, or if in the United States, call 1-800-829-4444 (8 am - 8 pm ET, Monday - Friday).