Zero+Cal

The N1913/1914A EPM Series power meters are able to perform Zero+Cal for the
U8480 USB thermocouple sensor via the front panel and SCPI. The Zero+Cal
softkey will be greyed out if no sensor is connected to that particular channel.
When the Zero+Cal softkey is pressed, zeroing and external calibration will be
triggered for the U8480 USB thermocouple sensor. The calibration type will not
be changed. The U8480 USB thermocouple sensor should be connected to the
reference calibrator before Zero+Cal is performed.

FDO Table Editing

When editing FDO table entries, the kHz softkey will be enabled when a U8480
USB thermocouple sensor is connected to a N1913/1914A EPM Series power
meter. The minumum entry for the frequency value is 0 kHz.
Figure 8-9
Keysight N1913/1914A User's Guide
Using U8480 Series USB Thermocouple Sensors
FDO table editing
8