Figure 6-5 Reference Calibration Factor Pop-Up Window - Keysight N1913 User Manual

Epm series power meters
Hide thumbs Also See for N1913:
Table of Contents

Advertisement

Procedure
1 Ensure that the power sensor is disconnected from any signal source.
2 Refer to the connection requirements in
3 Check the current reference calibration factor setting by pressing
Does this setting match the value for the sensor? (The power sensor reference
calibration factor can normally be found above the calibration factors table on the
power sensor body.)
4 To change the settings, press the channel
Figure 6-5
5 Press
Now, zero and calibrate the power meter/sensor combination as follows:
6 Press
7 Connect the power sensor to the POWER REF output.
8 Press
Keysight N1913/1914A User's Guide
is ready for connection to the POWER REF.
,
. The value is displayed under the channel
1 of 2 REF CFs
factor pop up window is displayed as shown in
keypad to enter the required value in the Ref Cal Factor pop-up menu.
Reference calibration factor pop-up window
to complete the entry.
%
and the channel
and the channel
Calibrating pop-up is then displayed.
Using N8480 Series Power Sensors
Table 6-2
and ensure that the sensor
. The reference calibration
REF CF
Figure
softkey. The Zeroing pop-up is displayed.
Zero
softkey to start the calibration routine. The
Cal
,
softkey.
Ref CF
6-5. Use the numeric
6
183