9
Maintenance
Trigger ignored
211
Indicates that <GET> or *TRG, or TRIG:IMM was received and recognized by the device but was ignored
because the power meter was not in the wait for trigger state.
Init ignored
213
Indicates that a request for a measurement initiation was ignored as the power meter was already initiated.
For example, INIT:CONT ON
INIT.
Trigger deadlock
214
TRIG:SOUR was set to HOLD or BUS and a READ? or MEASure? was attempted, expecting TRIG:SOUR to be
set to IMMediate.
Parameter error;Frequency list must be in ascending order.
220
Indicates that the frequencies entered using the MEMory:TABLe:FREQuency command are not in ascending
order.
Settings conflict
221
This message occurs under a variety of conflicting conditions. The following list gives a few examples of
where this error may occur:
If the READ? parameters do not match the current settings.
If you are in fast mode and attempting to switch on for example, averaging, duty cycle or limits.
Trying to clear a sensor calibration table when none is selected.
Data out of range
222
A numeric parameter value is outside the valid range for the command.
For example, SENS:FREQ 2 kHZ.
Illegal parameter value
224
A discrete parameter was received which was not a valid choice for the command. You may have used an
invalid parameter choice.
For example, TRIG:SOUR EXT.
Lists not same length
226
This occurs when SENSe:CORRection:CSET[1]|CSET2:STATe is set to ON and the frequency and calibration/
offset lists do not correspond in length.
Data corrupt or stale
230
This occurs when a FETC? is attempted and either a reset has been received or the power meter state has
changed such that the current measurement is invalidated (for example, a change of frequency setting or
triggering conditions).
Data corrupt or stale;Please zero and calibrate Channel A
230
When CAL[1|2]:RCAL is set to ON and the sensor currently connected to Channel A has not been zeroed and
calibrated, then any command which would normally return a measurement result (for example FETC?,
READ?, or MEAS?) will generate this error message.
228
Keysight N1913/1914A User's Guide