Section 6: SCPI command reference

:TRIGger:LOAD "ConfigList"

This command loads a predefined trigger model configuration that uses source and measure configuration lists.
Type
Command only
Usage
:TRIGger:LOAD "ConfigList", "<measureConfigList>", "<sourceConfigList>"
:TRIGger:LOAD "ConfigList", "<measureConfigList>", "<sourceConfigList>", <delay>
:TRIGger:LOAD "ConfigList", "<measureConfigList>", "<sourceConfigList>", <delay>,
"<bufferName>"
:TRIGger:LOAD "ConfigList", "<measureConfigList>", "<sourceConfigList>", <delay>,
"<bufferName>", <readingBlock>
<measureConfigList>
<sourceConfigList>
<delay>
<bufferName>
<readingBlock>
Details
This trigger model template incorporates a source configuration list and measure configuration list.
You must set up the configuration lists before loading the trigger model.
You can also set a delay and change the reading buffer.
After selecting a trigger model template, you can view the trigger model blocks in a graphical format
by pressing the front-panel MENU key and under Trigger, selecting Configure. You can also add or
delete blocks and change trigger model settings from this screen. You can use the
TRIGger:BLOCk:LIST? command to view the trigger model blocks in a list format.
This command replaces the TRIGger:LOAD:CONFiguration:LIST command, which is
deprecated.
6-222
Model 2461 Interactive SourceMeter® Instrument Reference Manual
Affected by
Not applicable
A string that contains the name of the measurement configuration list to
use
A string that contains the name of the source configuration list to use
The delay time before each measurement (167 ns to 10 ks); default is 0 for
no delay
A string that indicates the reading buffer; the default buffers (defbuffer1
or defbuffer2) or the name of a user-defined buffer; if no buffer is
specified, defbuffer1 is used
Define a measure or digitize block for the trigger model; options are:
ACTive: Add a measure or digitize block to the trigger model based
on the active function; if no option defined, ACTive is used
MEASure: Adds a measure block to the trigger model
DIGitize: Adds a digitize block to the trigger model
Where saved
Not applicable
Default value
Not applicable