Section 8: TSP command reference
Example 2
print(trigger.model.getblocklist())
Query the configuration of the blocks in the trigger model.
Output:
4) SOURCE_OUTPUT
5) SOURCE_PULSE_OUTPUT
Also see
trigger.model.setblock() — trigger.BLOCK_SOURCE_OUTPUT
trigger.model.setblock() — trigger.BLOCK_WAIT
This function defines a trigger model block that waits for an event before allowing the trigger model to continue.
Type
TSP-Link accessible
Function
Yes
Usage
trigger.model.setblock(blockNumber, trigger.BLOCK_WAIT, event)
trigger.model.setblock(blockNumber, trigger.BLOCK_WAIT, event, clear)
trigger.model.setblock(blockNumber, trigger.BLOCK_WAIT, event, clear, logic, event)
trigger.model.setblock(blockNumber, trigger.BLOCK_WAIT, event, clear, logic, event,
event)
blockNumber
event
clear
logic
Details
You can use the wait block to synchronize measurements with other instruments and devices.
You can set the instrument to wait for the following events:
Front-panel TRIGGER key press
Notify (only available when using remote commands)
Command interface trigger
Digital input/output signals, such as DIGIO and TSP-Link
LAN
Blender
Timer
Source limit condition
8-298
Model 2461 Interactive SourceMeter® Instrument Reference Manual
OUTPUT: ON
PULSE OUTPUT: ON
Affected by
Restore configuration
Instrument reset
Power cycle
The sequence of the block in the trigger model
The event that must occur before the trigger block allows trigger execution
to continue (see Details)
To clear previously detected trigger events when entering the wait block:
trigger.CLEAR_ENTER
To immediately act on any previously detected triggers and not clear them
(default): trigger.CLEAR_NEVER
If each event must occur before the trigger model continues:
trigger.WAIT_AND
If at least one of the events must occur before the trigger model continues:
trigger.WAIT_OR
(on page 8-296)
Where saved
Configuration script
Default value
Not applicable