Section 6: SCPI command reference

[:SENSe[1]]:CONFiguration:LIST:CREate

This command creates an empty measure configuration list.
Type
Command only
Usage
[:SENSe[1]]:CONFiguration:LIST:CREate "<name>"
<name>
Details
This command creates an empty configuration list. To add configuration indexes to this list, you need
to use the store command.
Configuration lists are not saved when the instrument is turned off. To save a configuration list, use a
saved setup to store the instrument settings, which include defined configuration lists.
Example
:SENS:CONF:LIST:CRE "MyMeasList"
Creates a measure configuration list named MyMeasList.
Also see
*SAV
(on page 6-15)
Configuration lists
[:SENSe[1]]:CONFiguration:LIST:STORe

[:SENSe[1]]:CONFiguration:LIST:DELete

This command deletes a measure configuration list.
Type
Command only
Usage
[:SENSe[1]]:CONFiguration:LIST:DELete "<name>"
[:SENSe[1]]:CONFiguration:LIST:DELete "<name>", <index>
<name>
<index>
6-74
Model 2461 Interactive SourceMeter® Instrument Reference Manual
Affected by
Recall settings
Instrument reset
Power cycle
A string that represents the name of a measure configuration list
(on page 3-30)
(on page 6-78)
Affected by
Not applicable
A string that represents the name of a measure configuration list
A number that defines a specific configuration index in the configuration list
Where saved
Save settings
Where saved
Not applicable
Default value
Not applicable
Default value
Not applicable
2461-901-01 A/November 2015