Smu.digitize.range - Keithley 2461 Reference Manual

1kw pulse mode interactive sourcemeter instrument
Table of Contents

Advertisement

Section 8: TSP command reference
smu.digitize.limit[Y].fail
smu.digitize.limit[Y].high.value

smu.digitize.range

This attribute determines the positive full-scale measure range for digitizer.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
value = smu.digitize.range
smu.digitize.range = value
value
Details
When you assign a range value, the instrument selects a fixed range that is large enough to measure
the assigned value. The instrument selects the best range for measuring the maximum expected
value.
For example, for digitize current measurements, if you expect a reading of approximately 9 mA, set
the range to 9 mA to select the 10 mA range.
When you read this setting, you see the positive full-scale value of the measurement range that the
instrument is presently using.
Example
smu.digitize.func = smu.FUNC_DIGITIZE_VOLTAGE
smu.digitize.range = 90
Also see
None
8-118
Model 2461 Interactive SourceMeter® Instrument Reference Manual
(on page 8-113)
(on page 8-115)
Affected by
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Set to the maximum expected value to be measured:
Current: 1 µA to 10 A
Voltage: 0.2 V to 100 V
Where saved
Configuration script
Measure configuration list
Set the range to 90 V, which selects the
100 V range
Default value
Current: 100 mA
Voltage: 7 V
2461-901-01 A/November 2015