Keithley 2461 Reference Manual page 1056

1kw pulse mode interactive sourcemeter instrument
Table of Contents

Advertisement

Appendix C: Status model
Event numbers
You can decide what events to use to set and clear the bits in the Operation Event Register and
Questionable Event Register. The following table lists the event numbers and descriptions.
This table lists frequently used event numbers. This table does not list all event numbers.
Event number
2728
2730
2731
2732
2733
2734
2735
2736
2737
2738
2739
2740
2741
2742
2743
2744
2745
2771
2774
2775
2776
2777
2778
4916
4917
5080
5081
C-10
Model 2461 Interactive SourceMeter® Instrument Reference Manual
Event description for user
Trigger model was running, but then aborted by user action.
The trigger model was running, but has failed and returned to the idle possibly due to an
error in settings.
Trigger model was idle, but is now running.
Trigger model was running, but completed successfully, and is now in idle.
The trigger model has ended in specified block number due to the specified condition.
A trigger model block has logged a user-defined informational event message.
A trigger model block has logged a user-defined informational event message.
A trigger model block has logged a user-defined informational event message.
A trigger model block has logged a user-defined informational event message.
A trigger model block has logged a user-defined warning event message.
A trigger model block has logged a user-defined warning event message.
A trigger model block has logged a user-defined warning event message.
A trigger model block has logged a user-defined warning event message.
A trigger model block has logged a user-defined error event message.
A trigger model block has logged a user-defined error event message.
A trigger model block has logged a user-defined error event message.
A trigger model block has logged a user-defined error event message.
The instrument is nearing an internal temperature where corrective action will be
necessary. The next limit will automatically turn off source outputs or limit other
functionality.
The instrument is experiencing an internal temperature malfunction since it is unable to
supply readings and requires immediate service.
An internal malfunction had cleared itself and the instrument is operating normally.
The instrument is operating at normal temperature levels and fully functional.
The instrument temperature has reached a level in which source output and other
functionality will not be available.
The instrument temperature has fallen to a level in which source output and other
functionality is now available.
The specified reading buffer is cleared (empty).
The specified reading buffer is full.
The source output has gone over the source limit level and is being limited.
The source output has returned below the source limit level and is no longer being
limited.