The trigger model states are:
Example
:TRIG:STAT?
Also see
None
:TRIGger:TIMer<n>:CLEar
This command clears the timer event detector and overrun indicator for the specified trigger timer number.
Type
Command only
Usage
:TRIGger:TIMer<n>:CLEar
<n>
Details
This command sets the timer event detector to the undetected state and resets the overrun indicator.
Example
:TRIG:TIM1:CLEar
Also see
:TRIGger:TIMer<n>:COUNt
:TRIGger:TIMer<n>:STARt:OVERrun?
2461-901-01 A/November 2015
Idle: The trigger model is stopped.
Running: The trigger model is running.
Waiting: The trigger model has been in the same wait block for more than 100 ms.
Empty: The trigger model is selected, but no blocks are defined.
Building: Blocks have been added.
Failed: The trigger model is stopped because of an error.
Aborting: The trigger model is stopping because of a user request.
Aborted: The trigger model is stopped because of a user request.
Affected by
Not applicable
(on page 6-236)
Where saved
Not applicable
Trigger timer number (1 to 4)
(on page 6-240)
Section 6: SCPI command reference
An example output if the trigger model is
inactive and ended at block 9 is:
IDLE;IDLE;9
Default value
Not applicable
Clears trigger timer 1.